Adequate space, appropriate equipment and qualified personnel are required to establish a molecular pathology laboratory. While the specifics of the requirements may vary depending on the spectrum of the tests that will be performed, there are several basic criteria that need to be fulfilled for standardization. In this paper, the criteria required to establish a molecular pathology laboratory will be reviewed.
Required Physical Conditions
One of the most important points that should be taken
into consideration while designing a molecular pathology
laboratory is to create a plan to prevent contamination.
Polymerase chain reaction (PCR)-based methods are
especially susceptible to contamination [4]. To obtain
a large number of copies from a very small amount of
the target sequence with the PCR method provides an
important diagnostic advantage, but this ability also leads to
false results in case of contamination. False positive results
may occur due to contamination from sample to sample,
transport of amplicons from the previous amplification
of the same target, cross-contamination of different
reactions prepared simultaneously, and contamination
of the reagents with DNA templates [5]. In real-time
PCR methods; when the PCR reactions are finished with
fluorescence-based detection techniques, the analysis is
also completed at the same time. Since the PCR products
do not need to be reprocessed, the reaction tubes or closed
plates are not opened and the amplicon transport does not
occur [4]. The laboratories using real-time PCR methods
therefore have less risk of contamination [5].
Main procedures performed in a molecular pathology laboratory using PCR-based methods are pre-PCR procedures (sample preparation, PCR preparation) and post-PCR procedures (performing PCR and post-PCR analysis) [4]. It is critical to perform these operations in areas separated as clean and dirty. The clean area represents the area where all pre-PCR procedures (such as microdissection, DNA/RNA extraction, PCR preparation) are performed, and the dirty area represents the area where all post-PCR products (amplicons) are processed. The staff and researchers should keep all reagents, materials and equipment used in these areas separate at all times and never move them back from the dirty area to the clean area [6]. Contamination is significantly reduced by physical separation of the clean and dirty areas and by doing pre-PCR and post-PCR activities in separate rooms. Therefore, planning at least two separate rooms is essential while designing a molecular pathology laboratory. It is recommended to perform sample preparation steps such as nucleic acid isolation in the pre-PCR laboratory, and to perform PCR reactions and other post-PCR procedures in the post-PCR laboratory [4]. However, if there is sufficient space, four separate rooms are recommended for the preparation of reagents, sample preparation, PCR step and post-PCR steps for an ideal molecular pathology laboratory. Each room must have its own equipment, protective clothing and consumables, and there should be no material/equipment transport between the rooms [7]. The requirements for laboratory design may vary according to the method used. For example, as mentioned previously, 3 rooms may be ideal in a laboratory where a real-time PCR method is applied, since post-PCR analysis is not necessary in the real-time PCR method [5].
Reagent preparation room is the room where reagent stocks are prepared and then divided into a certain number of small usable parts (aliquoted), and the reaction mixes are prepared. This room should be free of any biological materials such as DNA/RNA extracts, PCR products, etc. The sample separation room is where the nucleic acid isolation is performed and the isolated samples are added to the PCR reaction mixes [5]. This room is also called a low copy room, as the number of copies has not yet been amplified by PCR [8]. Ideally, it is recommended to perform the steps of nucleic acid isolation and addition of isolated samples to the PCR reaction mixes in separate rooms, but these two steps are usually performed in the same room but in different areas/compartments since most laboratories do not have sufficient space [5]. Preparation of the PCR reactions in a laminar flow biosafety cabinet ensures that the area remains clean [9]. The amplification (PCR) room is where PCR devices are located and the amplification steps are performed, and the post-PCR room is where the analysis of PCR products by gel electrophoresis, sequencing, nested-PCR, etc., methods are carried out. These two rooms constitute contaminated-dirty rooms and no equipment or materials used in these rooms should be used in other rooms [5]. These rooms are also called high copy rooms [8]. In the PCR applications such as real-time PCR where single-stage PCR reactions are sufficient and tubes with PCR products are not required to be opened, PCR devices can be placed in the post-PCR room. However, in the laboratories using PCR applications such as nested PCR, etc., where multiple PCR reaction steps are required and the tubes must be opened, PCR devices should be placed in a separate room/area [9]. In the amplification phase, the primary and secondary PCR steps (if any), should be separated according to the physical state of the laboratory, preferably in separate rooms. If this is not possible, they should be performed in separate compartments and on separate PCR devices [7]. Next-generation sequencing applications also include one or more PCR amplification steps, which are similarly recommended to be performed in separate rooms/areas [5]. Various recommendations about minimum room sizes can be found in international guidelines. For example, according to the field planning criteria of the United States of America Military Health System Pathology and Clinical Laboratories guide, the reagent preparation room should be at least 120 sq ft (approx. 33.5 m2) and the amplification room should be 240 sq ft (approx. 22.3 m2) in size [10]. The relevant guide published by the Republic of Turkey Ministry of Health is detailed below.
Workflow
The workflow in the molecular pathology laboratory must
be unidirectional from the clean area to the dirty area [7].
When laboratory personnel and researchers are required to
move from dirty rooms to clean rooms, laboratory coats,
gloves and all kinds of protective equipment should be
changed and hands should be washed. No material should
be carried from the dirty room to the clean room [5]. To
prevent the passage of personnel from the dirty room to
the clean room, it is appropriate to have separate personnel
working in each room or to perform pre-PCR and post-
PCR procedures on different days [9]. There are automated
molecular pathology platform systems that provide
automatic one-way workflow and isolate nucleic acid from
the sample, combine the isolated DNA with amplification
reagents, and perform the analysis, and their use is becoming
increasingly common [5,11]). The rooms and workflow
that should be present in an ideal molecular pathology
laboratory are shown in Figure 1. If all operations have to
be performed in a singl room, separate compartments/
benches are required for the reagent preparation, sample
preparation, PCR stage and post-PCR stages. The rule of
unidirectional workflow from the clean compartments
to the dirty compartments must be followed. If possible, sample preparation should be carried out in a laminar flow
biosafety cabinet including UV light. In the absence of
separate rooms, a timetable should be established in which
the pre-PCR and post-PCR steps are performed at different
times of the day [4].
Ventilation
Circulating air between pre- and post-PCR laboratories is
an important source of contamination in laboratories where
techniques detecting very small amounts of DNA/RNA are
used. Each laboratory should be ventilated separately and
the air pressure must be adjusted separately. At positive pressure, the air pressure inside the room is higher than
the air pressure outside the room, preventing the transport
of unwanted substances from outside. Negative pressure,
on the other hand, allows air to enter into the room and
prevents the migration of the air to the surrounding rooms/
laboratories. The doors must be kept closed to maintain
the negative pressure. There should be slight positive
pressure in the pre-PCR laboratory to prevent the entrance
of contaminated air from outside, while the post-PCR
laboratory should have a slight negative pressure to keep
the air in and thus to prevent the escape of amplicons from
the completed PCR samples. The ventilation of pre-PCR
and post-PCR laboratories should be opened to different
air channels and opened out from different locations [4].
Ultra-Violet (UV) Irradiation
UV rays that cause DNA damage are useful for eliminating
the contaminated DNA that may occur during addition
of the DNA template. UV light can therefore be used to
sterilize the pre-PCR laboratory. Since this method is
based on cross-linking with thymidine residues, the base
sequence of the target region plays a role in its success. In
addition, the hydration status of the DNA has a significant
effect on the UV resistance of the DNA. As dry-state DNA
is more resistant to UV light, UV light is less effective in
preventing contamination on dry laboratory surfaces
[9]. If UV light is going to be used on master mixes for
decontamination, care must be taken to ensure that no
dNTPs and enzymes are damaged in the UV light [8]. The
UV light source can be placed on the laboratory ceiling or
bench and can be activated by a device on the exit door
as the last person leaving the laboratory closes the outer
door. If UV lights are used, UV-induced ozone must be
removed by ventilation. Accumulation of deposits due
to the precipitation of oxidation products on the glass
of the bulb during radiation occurs and this reduces the
effectiveness of the UV system. These deposits should be
removed monthly and the performance of the UV bulbs
must be strictly monitored [4].
The physical conditions required in molecular laboratories in Turkey are defined by the Guideline for Physical Infrastructural Standards of Medical Laboratories applying Molecular Tests published by the Republic of Turkey Ministry of Health [12]. According to this guideline;
Molecular diagnostic laboratories should have at least two, preferably three rooms, each with a minimum area of 15 square meters, physically separated from each other to allow unidirectional workflow (from preamplification to postamplification) and preferably with separate ventilation systems.
These rooms are defined in parallel with those recommended in the literature and international guidelines, as a pre-amplification laboratory where sample acceptance and nucleic acid extraction are performed, an amplification laboratory where target amplification methods are applied, and a postamplification laboratory where analysis methods such as electrophoresis, DNA sequence analysis etc. take place after amplification.
In order to prevent contamination, the issues of providing a clean and preferably separate airflow, preparation and storage of all reagents and chemicals in their own areas, use of separate devices and materials for each laboratory area (freezers, refrigerators, cabinets, centrifuges, water baths, vortex mixers, pipettes, pencils, timers, all kinds of consumables, etc.) are pointed out.
It is emphasized that if only two rooms can be reserved for the molecular diagnosis laboratory, preamplification procedures and amplification / postamplification analyses should be performed in separate rooms.
For each laboratory; the issues of the presence of single piece floor covering without pores, the presence of hand washing sinks, the provision of temperature and humidity monitoring, the use of UV irradiation systems (on counter tops and / or room ceilings) during non-working hours, the presence of sufficient storage space, the presence of a sufficient number of grounded electrical outlets and uninterruptible power supplies (UPS, generator, etc.) for the laboratory devices, and the placement of laboratory equipment to allow unidirectional workflow are pointed out.
What to Do to Avoid Intermixing and Contamination of
Samples
Molecular laboratory tests are generally quite sensitive and
specific, providing very precise results [6]. Even so, false
positive or false negative results may sometimes occur.
Control mechanisms that include the verification of the
primer and probe sequences, checking and confirming
whether the test conditions are optimal, and the use of
negative controls [9] should be employed to reduce false
results.
While amplification is generally a part of the molecular diagnostic method, current nucleic acid amplification methods are very sensitive with the capability of detecting even a single molecule. Although this seems to be an advantage, it should be kept in mind that the contaminated DNA molecule may also be amplified, causing false positive results. Therefore, prevention of contamination must be a priority in molecular pathology laboratories [9].
Cross-contamination is one of the sources of error and contamination in the pathology laboratory and may occur at any stage of the tissue processing process, such as during macroscopic and/or microscopic evaluation, or during DNA extraction, and may cause false positive or negative results. Microorganisms (viruses, bacteria, etc.) may also be transferred from one case to another during these processes. Immunohistochemical staining with ABH blood group antibodies, microdissection, and microsatellite instability analysis can be performed to prevent and detect crosscontamination [2,13]. As processing of the samples from different patients occurs in the same area with recurrent use of several instruments (e.g., microtome blade, water bath) in pathology laboratories, precautions such as using a new blade for each sample, washing the blade with DNA decontamination solution, and/or sectioning of an empty paraffin block between samples (called the sandwich model) are used by various laboratories to reduce the risk of cross-contamination [13]. However, cross-contamination rates have been reported to be around 3% (0-8.8 %) despite these precautions [14].
The amplified DNA from the positive reactions in the previous test, when the reaction tubes are opened after amplification, is the source of contamination for the subsequent tests [9]. Also, amplification reactions are exposed to contamination from other patients samples and the target-containing plasmid [15]. Samples may contaminate the laboratory environment while pipetting and the risk of contamination increases if multiple samples are run together. Positive controls studied in the test are risk sources for contamination as well. Clothing, laboratory waste and/or uncleaned tables may contain contaminating nucleic acids [9].
In order to prevent and control contamination, the appropriate physical conditions, architectural structure and design, meticulous application of laboratory techniques and environmental control protocols, and the workflow plan are essential. Using separate areas or rooms for pre-amplification, amplification and postamplification stages with separate ventilation systems is an efficient way to prevent contamination, as discussed in detail in the Physical Conditions section above. The risk of contamination is lower in closed system devices. Only personnel in charge should be present in the test area. Every area/room should have its own equipment including laboratory coats and pipettes [4,6,15-17]. Minimum aerosolization while opening tubes is also necessary to prevent transport between samples [15].
Cleaning before and after each procedure should be carried out by using nucleic acid removers. For example, washing with 10% bleach that is freshly prepared and rinsing with 70% ethanol can be performed [16]. Thus, both biologically hazardous substances and nucleic acids that may be sources of contamination can be removed [15]. Adding enzymes such as uracil DNA glycosylase, which break down DNA, into the amplified DNA to exchange some or all of the thymidine with uracil in the reaction products can prevent contamination biochemically [18]. In addition, contamination can be prevented or reduced by discarding unopened tubes in the last stage, using pipettes with positive pressure displacements, not talking during tests such as PCR, and regular exposure of laboratory devices to UV radiation [4,6]. Aliquoting the reagents for each run is another precaution for the prevention of contamination. Patient samples and positive controls should be the last to be put into reaction tubes to reduce the risk of transport of the nucleic acids. If positive controls are going to be used, the lowest dilutions should be preferred [15].
Water or DNA-free buffers can be used as negative controls to detect and monitor contamination. The control tube should contain all materials for all stages of the test, like any other sample. A positive result detected in the negative control indicates the possibility of contamination [4,9]. The adequacy control of chemical sterilization (such as uracil glycosylase protocols) can be done by incorporating a small amount of amplicon into a negative control [15]. Surface and equipment contamination can be checked by swab samples from the laboratory surfaces where the test is carried out with damp filter papers, and a positive result in the swab sample indicates the presence of contamination. In addition, more than expected positivity rates of a given test may indicate contamination [4,9].
As RNA is more reactive than DNA and is vulnerable to RNAses that are present in all cells, prevention of RNAse contamination is very important in RNA-based molecular tests. RNases are resistant to metal chelating agents and can persist even after prolonged boiling or autoclaving. The most common sources of exogenous RNAse contamination are contaminated buffers and automatic pipettors. Also, all laboratory surfaces and glassware can be contaminated with RNAse from the laboratory personnels skin, hair, etc. Wearing gloves and changing them frequently during all stages of the test, use of separate laboratory equipment for RNA based tests, aliquoting small amounts of buffers, use of RNAse inhibitors (DEPC, etc.), and use of RNAase-free solutions and tubes are among the laboratory precautions to prevent RNAase contamination [19].
Preparation of a documented action plan in case of contamination is recommended. Most laboratories quarantine and/or destroy all contaminated reagents and consumables [15].
Equipment
Equipment and appliances may slightly differ between
molecular pathology laboratories based on their testing
profile. On the other hand, often a detailed inventory
list (Table I) is required to set up a molecular pathology
laboratory and provide standard testing. It should also be
noted here that equipment and consumables used in the
routine pathology setting must be available in a molecular
pathology laboratory, as well. It is also important to budget
for service contracts for maintenance and repair [6].
Unidirectional workflow should be taken into consideration
while organizing/placing the equipment (Figure 2). The list
of the appliances and equipment required in a molecular
pathology laboratory provided by Republic of Turkey
Ministry of Health is also a useful guide (Table II) [12].
Table I: The list of equipment and appliances required in a molecular pathology laboratory.
Table II: The list of appliances and equipment required in a molecular laboratory as set by the Republic of Turkey Ministry of Health [12].
Instructions for calibration and maintenance should be kept in the laboratory as written guidelines. Calibration guidelines must include the schedule for calibration (e.g., daily, monthly etc.), instructions describing the steps of the calibration procedure, calibration material specifications, preparation and storage conditions, troubleshooting and documentation methods, maintenance guidelines, the schedule for maintenance, instructions for performing maintenance, and troubleshooting guidelines [20].
Quality Assessment
Quality management is essential in all steps of pathology
evaluation (i.e., pre-analytical, analytical and postanalytical)
and a very important component in molecular
pathology practice. However, as the details of quality
management are beyond the scope of this review, only the
basic principles will be mentioned here.
There are several regulatory guidelines, including standard operating procedure manuals, to be followed to set-up and/ or manage a molecular pathology laboratory [12,17,20]. The presence of errors affecting the accuracy of the results in a molecular pathology laboratory should be checked regularly under the supervision of the pathologist (quality control; QC) to prevent or minimize erroneous reports and provide the confidence that quality requirements will be fulfilled (quality assurance; QA) [21].
Turn-around time and test result statistics should be checked and validated by using standard validation studies [22]. For internal quality assessment (IQA), the use of control materials is recommended (see previous sections for details).
In addition to the internal precautions mentioned in the previous sections, external quality assessment (EQA) must also be performed at given intervals for specific types of tests, as it is the most critical stage of quality management.
EQA, a measure of laboratory performance [23], has been shown to be helpful to improve molecular pathology laboratories [24], and EQA programs are the key elements of a laboratorys QA framework [25]. Regular participation in EQA is needed to verify and improve the quality of testing, as molecular pathology EQA schemes score the report and the test result [26,27]. As a part of an EQA scheme, participants receive test samples and their results are then reviewed to check for errors.
The reports are scored and the participant laboratories gain the opportunity to improve their service. Laboratories across Europe are also required to have accreditation [26]. Accreditation is a process in which an authorized independent body officially recognizes that the laboratory is competent to perform certain tasks, and may be considered the most effective system for QA as compliance with ISO standards is checked by accreditation bodies. Both accreditation and participation in EQA are recognized as effective and important tools to improve the accuracy and reliability of molecular testing [28].
CONFLICT of INTEREST
The authors declare no conflict of interest.
ACKNOWLEDGEMENT
We would like to thank to Professor Kutsal Yörükoğlu and
Professor Sülen Sarıoğlu for their guidance and valuable
contributions during the writing of this article.
FUNDING
No funding to declare.
1) van den Tweel JG, Taylor CR. A brief history of pathology:
Preface to a forthcoming series that highlights milestones in
the evolution of pathology as a discipline. Virchows Arch. 2010;
457:3-10.
2) Fassan M. Molecular diagnostics in pathology: Time for a nextgeneration
pathologist? Arch Pathol Lab Med. 2018; 142:313-20.
3) Uzun E, Sarioğlu S. Techniques for maximizing the performance
of molecular pathology testing: Responsibilities of all pathologists.
Turk Patoloji Derg. 2018;34:113-26.
4) Mifflin TE. Setting up a PCR laboratory. CSH Protoc. 2007;2007:
pdb.top14.
5) Standards Unit, Public Health England. UK Standards for
Microbiology Investigations: Good practice when performing
molecular amplification assays. (Accessed on January 28
2020). Available from: https://assets.publishing.service.gov.
uk/government/uploads/system/uploads/attachment_data/
file/682533/Q4i5.pdf
6) Hunt JL. Molecular pathology in anatomic pathology practice: A
review of basic principles. Arch Pathol Lab Med. 2008;132:248-60.
7) McDonagh S. Equipping and Establishing a PCR Laboratory. In:
2nd ed. Bartlett J, Stirling D, editors. PCR Protocols: Methods in
Molecular Biology. New Jersey: Humana Press; 2003 (226). 15-20.
8) Khan Z. Setup of a PCR Laboratory. In: Park DJ, editor. Methods
in Molecular Biology. New York: Springer Science+Business
Media; 2011. 3-14.
9) Lo YMD, Chan KCA. Setting Up a Polymerase Chain Reaction
Laboratory. In: Lo YMD, Chiu RWK, Chan KCA, editors. Clinical
Applications of PCR. Methods in Molecular Biology. New Jersey:
Humana Press; 2006 (336). 11-8.
10) Defense Health Agency Facilities. Chapter 530: Pathology and
Clinical Laboratory, DoD Space Planning Criteria. (Accessed on
January 28 2020). Available from: https://www.wbdg.org/FFC/
DOD/MHSSC/spaceplanning_healthfac_530_2017.pdf
11) Felder RA, Jackson KD, Walter AM. Process evaluation of an
open architecture real-time molecular laboratory platform. J Lab
Autom. 2014;19:468-73.
12) T.C. Sağlık Bakanlığı. Moleküler testler çalışılan tıbbi
laboratuvarların fiziksel altyapı standartları ile ilgili kılavuz.
(Accessed on January 28 2020). Available from: https://
dosyamerkez.saglik.gov.tr/Eklenti/2619,moltetaltyapikilavuzup
df.pdf?0.
13) Asor E, Stav MY, Simon E, Fahoum I, Sabo E, Ben-Izhak O,
Hershkovitz D. Risk for molecular contamination of tissue
samples evaluated for targeted anti-cancer therapy. PLoS One.
2017;12(3):e0173760.
14) Worsham MJ, Wolman SR, Zarbo RJ. Molecular approaches
to identification of tissue contamination in surgical pathology
sections. J Mol Diagn. 2001;3:11-5.
15) Association for Molecular Pathology statement.
Recommendations for in-house development and operation of
molecular diagnostic tests. Am J Clin Pathol 1999;111:449-463.
16) Prince AM, Andrus L. PCR: How to kill unwanted DNA.
Biotechniques. 1992;12:358-9.
17) Establishing Molecular Testing in Clinical Laboratory
Environments; Approved Guideline CLSI. CLSI document
MM19-A. Wayne, PA: Clinical and Laboratory Standards
Institute, 2011.
18) Standards Unit, Department for Evaluations, Standards and
Training Centre for Infections. Good laboratory practice when
performing molecular amplification assays. QSOP 38. (Accessed
on January 28 2020). Available from: https://www.genequantification.
de/good-laboratory-practise-qpcr-qsop38-2010.
pdf
19) Green MR, Sambrook J. How to win the battle with RNase. Cold
Spring Harb Protoc. 2019;2019(2).
20) Clinical and Laboratory Standards Institute (CLSI). Laboratory
Documents: Development and Control; Approved Guideline-
Fifth Edition. Pennsylvania: CLSI Institute; 2006.
21) International Organizational for Standardization. ISO 9000:2015.
International standards for quality management. Quality
management systems Fundamentals and vocabulary. Genève:
International Organization for Standardization, 2015.
22) College of American Pathologists. Molecular Pathology Checklist.(
Accessed on April 01 2020). Available from: http://webapps.
cap.org/apps/docs/education/OnlineCourseContent/2012/LAP/
Resources/Checklists/MOL09252012.pdf
23) World Health Organization, Laboratory Quality Management
System (LQMS) Training Toolkit. External Quality Assessment.
Module 10. (Accessed on April 01 2020) Available from: https://
www.who.int/ihr/training/laboratory_quality/10_b_eqa_
contents.pdf
24) Tack V, Schuuring E, Keppens C, Hart N, Pauwels P, van Krieken
H, Dequeker EMC. Accreditation, setting and experience as
indicators to assure quality in oncology biomarker testing
laboratories. Br J Cancer. 2018;119:605-14.
25) James D, Ames D, Lopez B, Still R, Simpson W, Twomey P.
External quality assessment: Best practice. J Clin Pathol. 2014;
67:651-5.
26) Cree IA, Deans Z, Ligtenberg MJ, Normanno N, Edsjö A, Rouleau
E, Solé F, Thunnissen E, Timens W, Schuuring E, Dequeker E,
Murray S, Dietel M, Groenen P, Van Krieken JH. Guidance for
laboratories performing molecular pathology for cancer patients.
J Clin Pathol. 2014; 67:923-31.